Technical Specifications

    lightSTUDIO-L
    lightSTUDIO-L IR
    lightSTUDIO-L 850/940 nm
    lightSTUDIO-AW
    Light sources
    10 x iQ-LED V2 | 41 SMD high-power LEDs
    5 x iQ-LED V2 | 41 SMD high-power LEDs | 10x iQ-LED-IR | 11 additional IR LED channels
    10 x iQ-LED V2 | 41 SMD high-power LEDs | 5 x iQ-LED 850 and 5 x iQ-LED 940 | 4 SMD high-power LEDs
    Image Engineering AW Light Source Technology (AW for Adaptive White)
    Intensity
    Intensity adjustable in 4000 steps per channel with 32 kHz PWM (switchable to 1000 steps with 128 kHz)
    Controlled via 4000 steps per channel and 32 kHz PWM (switchable to 1000 steps with 128 kHz)
    Intensity adjustable in 4000 steps per channel with 32 kHz PWM (switchable to 1000 steps with 128 kHz)
    2 lx – 1800 lx (1AW90 Head) 4 lx – 3600 lx (2AW90 Head) 1.5 lx – 1700 lx (1AW95) 3 lx – 3400 lx (2AW95)
    Spectral Range
    380-820 nm
    380-1050 nm
    380-820 nm
    -
    Predefined standard illuminants
    D50, D55, D65, D75, A, B, C, E - Create a Planckian spectral curve utilizing a selected temperature (1900 - 18,000 K)
    D50, D55, D65, D75, A, B, C, E - Planckian spectral curve by selected temperature (1900 – 18,000 K)
    D50, D55, D65, D75, A, B, C, E - Planckian spectral curve by selected temperature (1900 – 18,000 K)
    2700-6500 K (depending on variant)
    Illumination stability
    +/-1% when stabilized (2% after switching D illuminants in the first 5 seconds)
    iQ-LED V2: +/- 1% when stabilized (2% after switching D illuminants in the first 5 seconds) iQ-LED IR: +/- 2% when stabilized
    +/- 1% when stabilized (2% after switching D illuminants in the first 5 seconds)
    Stabilization time of <1.8 seconds per 1% intensity change Stabilization time is 0 seconds for a change in color temperature
    Response time (switch illuminant)
    < 50 ms
    < 50 ms
    < 50 ms
    Manual control via device’s control panel: < 5 ms Remote control from PC: < 120 ms
    Maximum/ Minimum illumination values
    Max. Up to 2000 lx for standard D illuminants Min. Down to 25 lx for standard D illuminants
    Max. Up to 1000 lux for standard D illuminants Min. Down to 25 lux for standard D illuminants
    Max. Up to 2000 lux for standard D illuminants Min. Down to 25 lux for standard D illuminants
    -